Safety Notices
Shock Hazard
The Cadex Battery Analyzer contains high-voltage circuits, and
can pose a shock hazard when the upper cover is removed. Do
not attempt to perform any service procedures on the analyzer
other than replacement of the external fuse (see Chapter 11,
Services and Upgrades, page 109).
To reduce the risk of electrical shock hazard, service batteries only when they
are removed and disconnected from their end-use equipment.
Use the analyzer only as specified in the documentation. Other uses may
impair the protection provided by the unit.
Use only a grounded AC outlet to power the analyzer.
Before attempting any internal service, remove all batteries from the battery
stations, turn the analyzer off and disconnect the power cord from the wall
socket. Wait a few minutes before opening the cover.
Replace fuses only with fuses of the same type and rating (see Replacing the
Primary (Input) Fuse, page 109).