• NO (default) sets the Cadex analyzer not to ask for the battery
C-code.
6. Press ENTER.
7. Press or to scroll to accept changes and press ENTER.
8. Press ESC twice to return to the global display.
Performing an Automatic OhmTest
Every time the Cadex analyzer runs a charge cycle, it can automatically perform an
OhmTest on the battery as well (except when the charge cycle is being performed
as part of a custom program). This test can be turned off.
Note Automatic OhmTest cannot be turned on or off using the following
programs: QuickTest, Q-Learn, Learn and Life-Cycle. OhmTest is included
in these programs and the settings cannot be changed.
To turn Automatic OhmTest on or off
1. From the global display, press MENU.
2. Press or to scroll to PROGRAM SETTINGS and then press ENTER.
3. Press or to scroll to AUTOMATIC OHMTEST and then press ENTER.
4. Press or to scroll to the required setting:
• YES (default) sets the OhmTest to run automatically before every
charge cycle (except during custom programs).
• NO turns off the automatic OhmTest.
5. Press ENTER.
6. Press or to scroll to YES to accept changes and press ENTER.
7. Press ESC twice to return to the global display.
Cadex Electronics Inc. C7000-C Series Battery Analyzers .