Optimizing Charge Time or Capacity
The end of charge conditions can be modified, (see Appendix B – Charge
Algorithms, page 148).
To select the charge optimization option
1. From the global display, press MENU.
2. Press or to scroll to PROGRAM SETTINGS and then press ENTER.
3. Press or to scroll to CHARGE OPTIMIZATION and then press ENTER.
4. Press or to move to between NICD AND NIMH OPTIMIZATION and
SLA AND LI OPTIMIZATION:
5. Press or to the required setting (see Appendix B – Change
Algorithms, page 148)
• TIME (default).
• CAPACITY.
6. Press ENTER.
7. Press or to scroll to YES to accept changes and press ENTER.
8. Press ESC twice to return to the global display.
Termination Condition on dT/dt
The Termination Condition on DT/DT parameter sets the temperature a battery can
reach before it will terminate or reduce the charge current. If a battery temperature
rises more than the DT/DT the charge current is either reduced or charge is
completed (see Appendix B – Charge Algorithms, page 148).
To select the charge optimization option
1. From the global display, press MENU.
2. Press or to scroll to PROGRAM SETTINGS and then press ENTER.
3. Press or to scroll to DT/DT TERMINATION and then press ENTER.
79