Duration: 3 minutes (user may repeat if unsuccessful).
To run the Boost program
1. Insert the Battery Adapter in a battery station.
2. Insert the battery in the Battery Adapter.
If the battery needs to be boosted, the global display for the station may show one
of two things:
i. It may continue to display the EMPTY message, indicating that the
Cadex analyzer has not recognized the battery.
ii. It may display a SHORTED (Code 121) message for the battery.
3. Press the station key and hold until a warning message appears to check
battery polarity (about 2 seconds).
4. If polarity is correct, press ENTER to start Boost. Depending on how the
analyzer is set up (see To verify C-code or enter Battery ID before battery
service begins, page 77), press ENTER again to accept the C-code settings
and start Boost.
5. The program runs for 3 minutes and provides a 100mA charge.
6. Press ESC any time during the Boost to terminate the program.
7. If the Boost is successful, the global display changes to the active program
name to show that the Cadex analyzer now recognizes the battery. If the
analyzer does not recognize the battery, remove and reinsert the battery and
the adapter.
8. If the Boost is unsuccessful, repeat this procedure until it is successful or until
you are convinced that the battery cannot be restored.
To edit the Boost settings
The Boost test program has no settings that can be edited. Boost current is fixed at
100mA for three minutes.