RS-232 baud rate: 300 to 115200 bps
Compliance: View compliance documents at
www.campbellsci.com/cs110
Lightning protection: Multi-stage transient protection on all external interfaces
Operating temperature range
Standard:
Extended (optional):
–25 to 50 °C
–40 to 85 °C
Operating relative
humidity range: 0 to 100% RH
Mounting: Vertical pipe with outer diameter (OD) of
1.91 to 6.35 cm ( 0.75 to 2.5 in)
Physical dimensions: 15.2 x 15.2 x 43.2 cm (6 x 6 x 17 in)
Weight:
4 kg (9 lb)
Construction:
l
Metallic ground connection to reciprocating shutter (no
wiping contact)
l
Brushless stepper motor
l
Powder-coated aluminum case
l
Teflon insulators
l
Electro-polished 316L stainless steel used for corrosion
protection of critical exposed metallic parts
6. Installation
Accurate electric field measurements are dependent on a site calibration factor, known as C
site
.
To maintain the validity of the site calibration factor, use the information in this section to select a
site and install the CS110. If these instructions are not or cannot be followed, new calibration may
be required (see Site correction (p. 22)).
6.1 Site selection 7
6.2 Installing the CS110 8
6.3 Power considerations 8
6.3.1 Connecting to AC power 9
6.4 Seal cable connectors 9