6. Connect the CS110 power cable to the CS110 POWER connector.
7. Route the cable down the pole and into the enclosure conduit.
8. Connect the black and clear wire to terminals on the PS150 or CH150 and red wire to the
+12 terminal.
9. Use cable ties to secure the cable to the pole.
10. Connect the green ground wire from the ground lug at the bottom of the enclosure to a
terminal on the PS150 or CH150.
J.3 Site calibration cautions
The greater the distance between the calibrator and the DUC, the greater the potential for error
in the resulting C
site
multiplier. Figure J-3 (p. 77) show the potential for error at separation
distances of 200 m, 15 m, 30 m and 10 m. An e
field
ratio (DUC/Calibrator) of 0.9 or 1.1 means that
the regression derived C
site
multiplier and resulting e
fields
will be off by 10%.