Technical Information [Chapter 5]
31
English
Electromagnetic Immunity (RF Conducted and Radiated)
The Capsule Neuron is intended for use in the electromagnetic environment specified in the table.
The customer or the user of the Capsule Neuron should make sure that it is used in such an
environment.
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans-
mitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Capsule Neuron is used exceeds the applicable RF compliance level above, the
Capsule Neuron should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the Capsule Neuron 2.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Immunity
Test
IEC 60601
test level
Compliance
level
Electromagnetic environment - Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 V Portable and mobile RF communications
equipment should be used no closer to any part of
the Capsule Neuron, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter. Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
. Interference may occur in the
vicinity of equipment marked with this symbol:
Radiated RF
IEC 61000-4-3
3 V/m 80
MHz to 2.5
GHz
3 V/m