Technical Information [Chapter 5]
32
English
Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the Capsule Neuron
The Capsule Neuron is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Capsule Neuron can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Capsule Neuron as recommended below,
according to the maximum output power of the communications equipment.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Rated maximum output
power of transmitter (W)
Separation distance in meters
according to frequency of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.70 3.70 7.37
100 11.70 11.70 23.30