EasyManua.ls Logo

Cellink INKREDIBLE - Page 61

Cellink INKREDIBLE
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
61
4. Adjust the X posion of printhead one unl it is right above the probing
sensor by following these steps:
-Select “Move Axis” and enter by pressing the control knob [8].
-Select “Move 0,1mm” and enter by pressing the control knob [8].
-Select “Move X” and enter by pressing the control knob [8].
-Jog the X-axis by turning the control knob [8] unl PH1 nozzle is in
the centre of the metal canlever of the sensor.
5. Adjust the Y posion of printhead one unl it is right above the probing
sensor by following these steps:
-Select “Move Axis” and enter by pressing the control knob [8].
-Select “Move 0,1mm” and enter by pressing the control knob [8].
-Select “Move Y” and enter by pressing the control knob [8].
-Jog the Y-axis by turning the control knob [8] unl PH1 nozzle is to
the round part of metal canlever of the sensor.
6. Return to the “Ulies Menu” and select the opon “Set Probe Posion” by
pressing the control knob [8].
7. Restart the INKREDIBLE 3D Bioprinter.
8.2.11 Pressure monitor not displaying correct pressure
The issue is most likely that the 0-level for pressure has been oset. To x this, do the
following:
1. Press and hold the buons on both sides of the middle buon on the
pressure display unl the pressure monitor’s 0-level has been reset.

Related product manuals