3. Use the following sequence to ne tune the hight of the printheads:
-Adjust set screw on the right side of printhead 2.
-In the “Ulies Menu”, select “Home printheads”.
-In the “Ulies Menu”, select “Raise printheads”.
-Control if the printheads is completely level. If not, repeat the sequence.
8.2.8 Display ickering or displaying random characters
The issue is most likely electrical interference from the air compressor. To x this, do the
following:
1. Turn o the INKREDIBLE 3D Bioprinter and the air compressor.
2. Disconnect the air compressor from the power outlet.
3. Connect the air compressor to a power outlet running on a dierent fuse.
4. Turn on the INKREDIBLE 3D Bioprinter and connue bioprinng.
8.2.9 Control knob is not responding when pressed
The issue is most likely that the knob was pushed all the way in, so it’s already in the
“click”-posion. To correct this, pull the knob gently towards you.
8.2.10 Printhead 1 is missing the probing sensor at the end of the “Home Axes”-
sequence
At the end of the “Home Axes” sequene, the nozzle in PH1 touches the probing sensor
aached to the printbed to calibrate the nozzle length. If the nozzle misses the probing
sensor, then the probing posion must be corrected to connue with the calibraon
process. Follow these steps to nd the probing posion for each axis:
1. Navigate to the “Ulies Menu” and enter by pressing the control knob [8].
2. Select HOME AXES and let the INKREDIBLE 3D Bioprinter carry out the
sequence. If printhead 1 does not trigger the probing sensor, trigger it
manually by hand by pressing the sensor 3 mes when the needle is about
to touch the sensor.
3. Adjust the Z posion of printhead one unl it is right above the probing
sensor by following these steps:
-In the “Ulies Menu”, select “Move Axis” and enter by pressing the
control knob [8].
-Turn the knob to job the Z-axis to posion Z+5.0 (The posion for each axis will
be wrien in the LCD display).