HITSTB Temp B High Alarm Test
Input: HITSTB
Returned: No response
Comments: Used to initiate the High Temp B Alarm Test. See the High
Temperature Alarm Test section for more information.
INITEE Restore All Defaults
Input: INITEE
Returned: No response
Comments: Used to restore all factory default settings.
LNSATP Set LN2 Saturation Temperature
Input: LNSATP ±XXX.X
Returned: No response
Comments: Used to set the LN2 saturation temperature for single point
calibration reference point.
Temperature Commands
BPTMP? Bypass Temp Query
Input: BPTMP?
Returned: ±XXX.X
Comments: Used to query the current bypass temp sensor reading.
HITA Set Temp A High Alarm
Input: HITA ±XXX.X
Returned: No response
Comments: Used to set the Temp A High Alarm value.
HITA? Temp A High Alarm Query
Input: HITA?
Returned: ±XXX.X
Comments: Used to query the current Temp A High Alarm value
HITAS? Temp A High Alarm Status Query
Input: HITAS?
Returned: 0 or 1
Comments: Used to query the Temp A High Alarm status where 0 = Off
and 1 = On.
HITAM Set Temp A High Alarm Mask
Input: HITAM 0 or HITAM 1
Returned: No response
Comments: Used to enable or disable the Temp A High Alarm where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
HITAM? Temp A High Alarm Mask Query
Input: HITAM?
Returned: 0 or 1
Comments: Used to query the Temp A High Alarm Mask status where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.