MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H • 152
HITB Set Temp B High Alarm
Input: HITB ±XXX.X
Returned: No response
Comments: Used to set the Temp B High Alarm value.
HITB? Temp B High Alarm Query
Input: HITB?
Returned: ±XXX.X
Comments: Used to query the current Temp B High Alarm value
HITBS? Temp B High Alarm Status Query
Input: HITBS?
Returned: 0 or 1
Comments: Used to query the Temp B High Alarm status where 0 = Off
and 1 = On.
HITBM Set Temp B High Alarm Mask
Input: HITBM 0 or HITBM 1
Returned: No response
Comments: Used to enable or disable the Temp B High Alarm where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
HITBM? Temp B High Alarm Mask Query
Input: HITBM?
Returned: 0 or 1
Comments: Used to query the Temp B High Alarm Mask status where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
LOTA Set Temp A Low Alarm
Input: LOTA ±XXX.X
Returned: No response
Comments: Used to set the Temp A Low Alarm value.
LOTA? Temp A Low Alarm Query
Input: LOTA?
Returned: ±XXX.X
Comments: Used to query the Temp A Low Alarm value.
LOTAS? Temp A Low Alarm Status Query
Input: LOTAS?
Returned: 0 or 1
Comments: Used to query the Temp A Low Alarm status where 0 = off and 1 =
on.
LOTAM Set Temp A Low Alarm Mask
Input: LOTAM 0 or LOTAM 1
Returned: No response
Comments: Used to enable or disable the Temp A Low Alarm where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
LOTAM? Temp A Low Alarm Mask Query
Input: LOTAM?
Returned: 0 or 1
Comments: Used to query the Temp A Low Alarm Mask status where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.