MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H • 153
LOTB Set Temp B Low Alarm
Input: LOTB ±XXX.X
Returned: No response
Comments: Used to set the Temp B Low Alarm value.
LOTB? Temp B Low Alarm Query
Input: LOTB?
Returned: ±XXX.X
Comments: Used to query the Temp B Low Alarm value.
LOTBS? Temp B Low Alarm Status Query
Input: LOTBS?
Returned: 0 or 1
Comments: Used to query the Temp B Low Alarm status where 0 = off and 1 =
on.
LOTBM Set Temp B Low Alarm Mask
Input: LOTBM 0 or LOTBM 1
Returned: No response
Comments: Used to enable or disable the Temp B Low Alarm where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
LOTBM? Temp B Low Alarm Mask Query
Input: LOTBM?
Returned: 0 or 1
Comments: Used to query the Temp B Low Alarm Mask status where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
TEMPA? Temp A Current Data Query
Input: TEMPA?
Returned: ±XXX.X
Comments: Used to query the current Temp A sensor reading.
TEMPB? Temp B Current Data Query
Input: TEMPB?
Returned: ±XXX.X
Comments: Used to query the current Temp B sensor reading.
TUNI Set Temp Units
Input: TUNI C, TUNI F, or TUNI K
Returned: No response
Comments: Used to set the Temp units where C = Celsius, F = Fahrenheit, and
K = Kelvin. Default setting is degrees Celsius.
TUNI? Temp Units Query
Input: TUNI?
Returned: C, F, or K
Comments: Used to query the Temp units where C = Celsius, F = Fahrenheit,
and K = Kelvin. Default setting is degrees Celsius.