EVNLOG? Event Log Record Query
Input: EVNLOG? N
Returned: XXXXX, MM/DD/YY, HH:MM, XXX.X, XXX.X, XXX.X, XXX.X
Comments: Used to query Event Record [n] where n = record number. String
returned: Date, Time, TempA, Temp B, LN2 Level, Liquid Usage, Event Codes.
LOGPER Set Event Log Interval
Input: LOGPER XXX
Returned: No response
Comments: Used to set the event log interval between 1 and 240 minutes.
Default setting is 240 minutes.
LOGPER? Event Log Interval Query
Input: LOGPER?
Returned: XXX
Comments: Used to query the current event log interval.
TIME Set the Time
Input: hh:mm:ss
Returned: No response
Comments: Used to set the controller Time in the format hh:mm:ss
(hour:minute:second).
TIME? Time Query
Input: TIME?
Returned: hh:mm:ss
Comments: Used to query the current time.
UNID Set Unit ID
Input: UNID XXXXX
Returned: No response
Comments: Used to set the controller 5-digit Unit ID. TEC 3000 Unit ID should
be between 1 and 200.
UNID? Unit ID Query
Input: UNID?
Returned: XXXXX
Comments: Used to query the controller Unit ID.