Digital Color Doppler Palm Ultrasound System
50
not lit When user turn on
the system.
connected, or may not be well
connected with the displayer.
Please contact with Sales
office, service department
and distributor.
Power indicating lamp is lit
When turn on the system,
but no images on displayer.
The restart time interval after
shutdown is too short
Wait 1 minute after power
off and then power on.
screen but no scanning
image.
Transmission frequency, gain
or STC control is not set
properly.
System is connected
improperly.
The system is in frozen status
frequency, gain or STC
control.
Ensure the system is
connecting correctly
Defreeze the system by
pressing the FREEZE
Image quality is abnormal
correct.
The image post-processing
setting is abnormal.
processing settings or set
it to default.
The system is not working
properly
Internal circuit protection
9.5 Service Responsibility
If users install, use and maintain the system fully according to CHISON’s installation manual, operation
manual and service manual, then SonoEye P2 main unit has a life time of 5 years,
The warranty of the system after ex-work is as the time in the warranty card.
The system is a precise electronic system. Standard maintenance must be performed by CHISON’s
authorized service engineer during the life time of the product.