Wound Component EST Scanner 19035/19035-M/19035-S
Quick Start Guide
<<< Troubleshooting >>>
14. Problems occurred
Problems occurred in the following situation are very dangerous. The
output terminal may still have high voltage output even the [STOP] key
is pressed; therefore, the user should be extremely careful when
dealing with it.
The DANGER LED indicator keeps on when [STOP] key is
pressed.
The DANGER LED indicator is on but the voltage meter has no
readings.
When the above situation occurs, shut down the power and unplug
the AC power cord immediately. Do not use the device again as
failure is awfully hazardous. Please send the hardware back to
Chroma or its distributor for repair service.
15. DANGER Indicator failure
When pressing the [START] key the voltage meter has readings but
the DANGER LED indicator is still off, it means the indicator may be
broken. Please power off the hardware and replace it with another
device, then send the broken one back to Chroma or its distributor for
repair service.
16. Be aware of the following when using the Scanner for long hour
in normal state.
If the high limit is set to 20.00mA (for hipot test), be aware of its
temperature change.
If the ambient temperature exceeds 40°C, stop using it
temporarily until the temperature drops to normal. Be sure to
check the room temperature before use.
17. There are four types of AC INPUT power applicable for this Scanner.
Switch the voltage selector on the rear panel to a proper voltage that
used locally. When plugging in the power cord, make sure the
inputted AC power is the same as the power range indicated on the
rear panel and the fuse is replaced with correct one. The table below
lists the voltage and the fuse it uses.
9