Programmable AC Source 61601/61602/61603/61604 Users Manual
4-4
Then press SHIFT, then to change to next step.
In above step C of V OUT AND MEAS. ACCURACY CALI., the user should not turn on the
load. Press
ENTER, then check the output voltage measured by DVM is about 15VAC.
This step is just to make sure the connection is correctly.
Then go to next step D. Press ENTER, check the output voltage measured by DVM is about
150VAC. Keyin the exact value measured by DVM, then press ENTER .
Then press SHIFT, then to change to next step.
In above step E of V OUT AND MEAS. ACCURACY CALI., the user should not turn on the
load. Press
ENTER, then check the output voltage measured by DVM is about 30VAC.
This step is just to make sure the connection is correctly.
Then go to next step F. Press ENTER, check the output voltage measured by DVM is about
300VAC. Key in the exact value measured by DVM, then press ENTER.
Step F is the final step of V OUT AND MEAS. ACCURACY CALI. Press PAGE/EXIT to exit
that page. Then display will show as below. Press ENTER to save the calibration results.
1. Users can press PAGE/EXIT to exit to the calibration choice page at
every step.
2. See the above display, if press PAGE/EXIT not to save the result, the
result of calibration still works, until turn the power off.
1. V OUT AND MEAS. ACCURACY CALI 150V RNG
C. ( ENTER ) THEN CHECK OUTPUT IS 15VAC
D. ( ENTER ) THEN KEYIN DVM MEAS. 150VAC
0.00_ V
1. V OUT AND MEAS. ACCURACY CALI 300V RNG
E. ( ENTER ) THEN CHECK OUTPUT IS 30VAC
F. ( ENTER ) THEN KEYIN DVM MEAS. 300VAC
0.00_ V
Press ( ENTER ) to save.
Press ( PAGE/EXIT ) not to save.