Remote Operation
7-7
*IDN? Return the AC source identification string
Return Parameter Chroma ATE 61600,123456,1.00,1.01,1.02
Chroma ATE : Company name
61600 : Model name
123456 : Serial number
1.00, 1.01,1.02
: Firmware version of display, waveform and remote
*RCL<n> Restore the values of the specific group which is previously stored in memory.
Parameter 1 - 3
*RST This command resets the AC source to the initial states. It’s better to wait
about 3 second to send the next command.
*SAV<n> Save the values into the specific group in memory.
Parameter 1 - 3
*SRE This command sets conditions of the Service Request Enabled Register. If
one or more of the enabled events of the Status Byte Register is set, the MSS
and RQS of Status Byte Register are set too.
*SRE? This query returns the Service Request Enabled Register.
*STB? This query returns the Status Byte Register.
Bit configuration of Status Byte Register
Bit Position
7
6
5
4
3
2
1
0
Condition
- -
MSS
RQS
ESB
MAV
QUES
- -
- -
- -
ESB = event status byte summary
QES = questionable status summary
RQS = request for service
MSS = master status summary
MAV = message available
*TST? This command queries the self-test result of the AC source.
7.6.2 Instrument Command Dictionary
The commands are listed in alphabetical order. Commands followed by question marks (?)
take only the query forms. When commands take both the command and query forms, they
are noted in the query syntax descriptions.