EasyManua.ls Logo

Chroma 61601 - SOURCE Sub-system

Chroma 61601
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Operation
7-11
Description : This command clears the latch that disables the output when an
overcurrent (OC), overtemperature (OT), overpower (OP) or
remote inhibit (RI) is detected. All conditions which have
generated the fault must be removed before the latch is cleared.
Query Syntax : None
Parameters : None
Return Parameters : None
7.6.2.3 SOURCE Sub-system
[SOURce :]
CURRent
: LIMit
: DELay
: INRush
: STARt
: INTerval
FREQency
[: {CW | IMMediate}]
VOLTage
[: LEVel][: IMMediate][:AMPLitude]
: AC
: DC
: LIMit
: AC
: DC
: PLUS
: MINus
: RANGe
[SOURce :] CURRent : LIMit
Description : This command sets the rms current limit of the AC source for
software protection.
Query Syntax : [SOURce :] CURRent : LIMit?
Parameters : <NR2>, valid range: 0.00 ~ maximum current spec. of the
specific model. (unit: A)
Return Parameters : <NR2>
[SOURce :] CURRent : DELay
Description : This command sets the delay time for triggering over current
protection.
Query Syntax : [SOURce :] CURRent : DELay?
Parameters : <NR2>, valid range: 0.0 ~ 5.0 (unit: 0.5 second)
Return Parameters : <NR2>
[SOURce :] CURRent : INRUsh : STARt
Description : This command sets the start time of the inrush current
measurement.
Query Syntax : [SOURce :] CURRent : INRUsh : STARt?
Parameters : <NR2>, valid range: 0.0 ~ 999.9 (unit: ms)
Return Parameters : <NR2>

Table of Contents

Other manuals for Chroma 61601

Related product manuals