Programmable DC Power Supply 62000P Series Operating & Programming Manual
4. Press
EXIT
” to return to Program PAGE (Figure 4-1).
See section 3.2 for detail description of settings.
4.2.5 Setting Voltage Slew_rate
1. Use
”, “
” keys to move the cursor to the column to be set as Figure 4-9 (5)
shows.
2. Use the numeric (
0
~
1
) keys or “Rotary” (
) knob to set the SEQ voltage conversion
slew rate.
3. Press
ENTER
” to confirm.
4. Press
EXIT
” to return to Program PAGE (Figure 4-1).
See section 3.2.2.3 for detail description of settings.
4.2.6 Setting Current Slew_rate
1. Use
”, “
” keys to move the cursor to the column to be set as Figure 4-9 (6)
shows.
2. Use the numeric (
0
~
1
) keys or “Rotary” (
) knob to set the SEQ current conversion
slew rate.
3. Press
ENTER
” to confirm.
4. Press
EXIT
” to return to Program PAGE (Figure 4-1).
See section 3.2.2.4 for detail description of settings.
4.2.7 Setting TTL OUT
1. Use
”, “
” keys to move the cursor to the column to be set as Figure 4-9 (7)
shows.
2. Use the numeric (
0
~
1
) keys or “Rotary” (
) knob to set the SEQ digital output state.
3. Press
ENTER
” to confirm.
4. Press
EXIT
” to return to Program PAGE (Figure 4-1).
See section 3.2.2.5 for detail description of settings.
4-14