3.7. SETTINGS
List and description of the parameters in the SETTINGS menu (default value in bold).
These parameters are common to all photometers types except when mentioned.
Auto: Enable or disable automatic measurement. The internal scheduler is
executed automatically [OFF; ON]: OFF.
Moon: Enable or disable moon measurement. The moon menus are invisible
in the SCENARIO menu if this setting is OFF. [OFF; ON]: OFF.
P. Moon: Period between two Group L scenarios. [2min; 15min]: 3min.
Country: Country identification. [0; 255]: 0.
District: District identification. [0; 255]: 0.
Number: Number identification. [0; 9999]: 0.
Latitude: Latitude of the current electronic box. [-90.00°; +90.00°]: 0.0000°.
Longitude: Longitude of the current electronic box. [-180.00°; +180.00°]:
0.0000°.
Altitude: Altitude of the current electronic box. [0m; 6000m]: 0m.
Turbo: Enable or disable the Turbo mode. [OFF; ON]: OFF.
P. Turbo: Period between two Group A scenarios. [2min; 6min]: 3min.
Cloud: Enable cloud mode. [OFF; ON]: OFF.
P. Bclsun: Bclsun scenario period. [30s; 360s]: 30min.
Bclsky: Amount of SKY in a BCLSKY scenario. [1; 1000]: 10.
P.bclsky: Delay between SKY in a BCLKSY scenario. [00:00:20; 00:06:00]:
[00:00:20].
AZ offset: Azimuth robot offset. [-180°; +180°]: .
ZN offset: Zenith robot offset. [-180°; +180°]: .
RG: Enable external Global Radiation measurement. [OFF; ON]: OFF.
RG sens: Global radiation sensitivity (µV/W/m²). [5.00 ; 50.00] : 10.00 ;
[DCP]
o See chapter 2.3.4.2 for detailed information.
[GSM]
o GSM Roaming: allow modem roaming for GSM/GPRS communication.
[OFF; ON]: OFF.
o [CSD]
CSD: Enable CSD communication. [OFF; ON]: OFF.
H0: Enable CSD communication between 0h00 and 0h59. [OFF;
ON]: OFF.
H1: Enable CSD communication between 1h00 and 1h59. [OFF;
ON]: OFF.

Table of Contents

Other manuals for Cimel CE318-T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cimel CE318-T and is the answer not in the manual?

Cimel CE318-T Specifications

General IconGeneral
BrandCimel
ModelCE318-T
CategoryMeasuring Instruments
LanguageEnglish