Cole Palmer 4-9
Section 4 Operation
FluidsType is used to identify the type of uid used. The controller uses the uid type to automati-
cally set certain operating parameters.
PumpAdjustment is used to review/set the
desired pump speed.
Water
EG-Water
PG-Water
Other
Menu
Low
High
Menu
2. With the desired uid selected press , or use
the Menu window, to save and return to
the previous display.
1. With FluidType highlighted, press to
display the list of acceptable uids.
Highlight the desired uid and press to
select it.
1. With PumpAdjustment highlighted, press
to display the speeds.
Highlight the desired speed and press to
select it.
Fluid system limits High °C Low °C
All units:
Water +98 +2
EG-Water +103 -33
PG-Water +103 -33
Controller 1 units:
Other +103 -33
Controller 2 units:
Other +103 -33
SIL 100 +75 -30
SIL 180 +150 -30
SIL 300 +150 +80
SYNTH 60 +45 -30
SYNTH 200 +150 +30
SYNTH 260 +150 +45
Controller 3 units:
Other +200 -90
SIL 100 +75 -75
SIL 180 +200 -40
SIL 300 +200 +80
SYNTH 60 +45 -50
SYNTH 200 +200 +30
SYNTH 260 +200 +45
NOTE The range is also limited by the bath
temperature range, see Section 2.