EasyManua.ls Logo

Cole Parmer POLYSTAT - Page 29

Cole Parmer POLYSTAT
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cole Palmer 4-9
Section 4 Operation
FluidsType is used to identify the type of uid used. The controller uses the uid type to automati-
cally set certain operating parameters.
PumpAdjustment is used to review/set the
desired pump speed.
Water
EG-Water
PG-Water
Other
Menu
Low
High
Menu
2. With the desired uid selected press , or use
the Menu window, to save and return to
the previous display.
1. With FluidType highlighted, press to
display the list of acceptable uids.
Highlight the desired uid and press to
select it.
1. With PumpAdjustment highlighted, press
to display the speeds.
Highlight the desired speed and press to
select it.
Fluid system limits High °C Low °C
All units:
Water +98 +2
EG-Water +103 -33
PG-Water +103 -33
Controller 1 units:
Other +103 -33
Controller 2 units:
Other +103 -33
SIL 100 +75 -30
SIL 180 +150 -30
SIL 300 +150 +80
SYNTH 60 +45 -30
SYNTH 200 +150 +30
SYNTH 260 +150 +45
Controller 3 units:
Other +200 -90
SIL 100 +75 -75
SIL 180 +200 -40
SIL 300 +200 +80
SYNTH 60 +45 -50
SYNTH 200 +200 +30
SYNTH 260 +200 +45
NOTE The range is also limited by the bath
temperature range, see Section 2.

Related product manuals