Warning: the quality of the EMG signal acquired by the Wave Plus system is linked to the quality of
the contact between electrode and skin; to obtain best results:
1. Use pre-gelled electrodes certified for a medical use and complying with the 93/42/CEE
regulation;
2. Do not use pre-gelled electrodes beyond the expiration date or with dry conductive gel;
3. Do not re-use the same electrode;
4. Do not use the electrode after having already applied it;
5. For long data acquisition periods, check the electrodes adhesive;
6. Apply electrodes only on undamaged skin and verify that their removal does not cause any
damage.
4.5 Electrodes battery recharge
To recharge the electrode batteries, put the electrode in the appropriate slot in the recharging module;
then power on the recharging module and wait for the signal that indicates the end of the recharging pe-
riod.
The jack socket for the power supply is accessible on the left side of the briefcase.
The charging module is equipped with an eight hours timer, after which the charge is automati-
cally stopped. This allows the unlimited permanence of electrodes into their respective hous-
ings, without the risk of overload the batteries.
To reset the timer and start a new charge cycle, you must disconnect the power jack and then
reconnect it.
4.5.1 LED indicator of the charging module
The operation of the charging module is indicated by a bicolor LED placed in the center of the unit;
alerts are:
1)Led Orange: the charger is on and the electrodes placed in the charging slots are subject to
charge;
2)LED green: the unit has completed the eight-hour charge cycle.
It is recommended to leave the case open during the recharge of the electrodes
units.