EasyManua.ls Logo

COMETA WAVE PLUS - Electrode Battery Replacement

COMETA WAVE PLUS
37 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WAVE PLUS - USER MANUAL
COD. WPMEMG
REV 01
02/11/2011
DATA PRIMA EMISSIONE
Pag. 26 di 37
Wave Plus system does not contain parts repairable by the user with the exception of the batteries; for
standard maintenance see the following chapters.
5.2 Replacement of electrode battery
After about 300 charging/discharging cycles, the battery could reduce its electric capacity that leads to a
shorter time of use of the electrodes. To re-enable the normal efficiency, replace the battery module as
follow:
1. Open the
probe by
gently push-
ing as shown
in the pic-
ture. The
probe will
slide open
separating in
two parts
(lower and
upper).
2. Take the
lower part of
the probe, the
one containing
the battery. As
shown in the
picture, the
battery is
linked to the
probe with a
small connec-
tor.