Dichiarazione del produttore - Immunità elettromagnetica II (Tabella 4, EN 60601-1-2)
Il dispositivo è adatto per l'uso in un ambiente elettromagnetico specico. Il paziente o il famigliare che utilizzano il
dispositivo devono assicurarsi che esso venga utilizzato in un ambiente elettromagnetico come descritto di seguito.
Manufacturer’s declaration - Recommended separation distances between portable and mobile RF
communication equipment and the product (Table 206, EN 60601-1-2)
The device is intended to operate in an electromagnetic environment where radio frequency radiated disturbances
are under control. The patient or family member using the device can help prevent electromagnetic interference by
ensuring a minimum distance between mobile and portable RF communication devices (transmitters) and the device as
recommended below, in relation to the maximum output power of radio communication equipment.
Immunity Test
IEC Test level
60601
Level of
compliance
Electromagnetic Environment (distance between
transmitter and device)
Radiated RF 10 V/m
80 MHz – 2,7 GHz
80% AM to 1 kHz
10 V/m
80 MHz – 2,7 GHz
80% AM to 1 kHz
d = 0,35 ·√P from 80 MHz to 800 MHz
d = 0,7 ·√P from 800 MHz to 1 GHz
d = 2,33 ·√ P from 1 GHz to 2,7 GHz
where P is the maximum output power rating in W
according to the transmitter’s manufacturer and
represents the separation distance in metres.
Potenza di uscita massima del
Trasmettitore specicata (W)
Distanza di separazione alla frequenza del trasmettitore (m)
Da 15 kHz a
80MHz
Da 80MHz a
800MHz
Da 800MHz
a 1GHz
Da 1GHz a
2,5GHz
0,01
0,1
1
10
100
0,12
0,04
0,07 0,23
0,11 0,22 0,74
0,35 0,70 2,33
1,11 2,21 7,38
3,50
7,00 23,33
0,37
1,17
3,69