Balancing
Rotating mass imbalance problems are analyzed and corrected
onsite. A rigid rotor can be balanced in one or two planes. The
following steps are involved:
An initial measurement of the imbalance is taken.
A trial weight is attached to the balancing plane and another
measurement is taken. For dual plane balancing, the same is
done for the second plane also.
Correction weights are attached to the balancing planes as
recommended by the instrument.
A measurement is taken in each plane to confirm that the
rotor is balanced. Any residual imbalance can be removed
via trim balancing.
See Balancing Rotors (page 123) and Balancing Walk-through (page
145) for more information.