3029-005B Section 5: Plumbing
Concordia Integra Technical Support Manual 5-3
2500-255B
Water Supply
Line Pressure
Line pressure must be between 30-100psi at 25gph. Adequate
line pressure is needed to fill both the hot water and steam
tanks.
Check Valve
The check valve is a back-flow prevention device which inhibits
reverse water flow into the water supply. Check to ensure the
check valve meets state and local health codes. Some local
codes may require a different type of back-flow prevention
device.
Water Filtration System
The Scalex® Water Treatment System includes a carbon filter
cartridge and a water softener cartridge.
Reverse Osmosis Water System
A fresh water bypass is required for sites with a reverse
osmosis filter system. The machine water level sensors are
inoperative when used with this system.