Alarm Function
Alarms are classified into two categories: technical alarm and physiological alarm.
Physiological Alarm
User can press 【MENU】 button to enter system menu, select 【ALARM SETUP】 item to enter its
interface, and then set alarm on-off and the high and low alarm limits, when blood pressure is higher than
the high limit or lower than the low limit, the physiological alarm will occur.
Press any button to cancel the alarm in physiological alarm state and this method brings no affection to the
next alarm. To close alarm in【ALARM SETUP】will disable the alarm function until switch alarm on.
Technical Alarm
When power is about to exhaust and alarm is on , then the alarm will occur. This alarm can not be
cancelled unless being closed or the power replaced.
SPO2 Measurement Function(Optional)
Please pay attention to:
Warning
◎ Pulse oximeter can overestimate the SPO2 value in the presence of Hb-CO, Met-Hb or
dye dilution chemicals.
◎ ES (Electrosurgery) equipment wire and SpO2 cable must not be tangled up.
◎ Do not put the sensor on extremities with arterial catheter or venous syringe.
◎ Do not perform SPO2 measuring and NIBP measuring on the same arm at one time, because
obstruction of blood flow during NIBP measuring may adversely affect the reading of
SPO2 value.
Note
◎ Make sure the nail covers the light window.
◎ the wire should be on the backside of the hand.
◎ SPO
2
value is always displayed in the fixed place.