Specification
Name Blood Pressure Monitor
Model CONTEC08A
Harmful liquid proof
degree
Ordinary equipment (sealed equipment without liquid proof)
Display 2.8'' color LCD Display
NIBP Specifications
Measurement Method Oscillometric method
Working modes Manual, Automatic
Measurement Range
Pressure:0~290mmHg(0~36.0kPa)
Pulse:40~240/min
Resolution
Pressure 1mmHg
Measurement Accuracy
Accuracy
±3mmHg
Error
The BP Value of the device is equivalence with the measurement
value of Stethoscopy . The error meets all the conditions in the
YY0667-2008.
Operating
Temperature/
Humidity
+5ºC~40 ºC . 15%RH~80%RH
Transport and Storage
Temperature/Humidit
y
-20ºC~+55ºC . ≤95%RH
Atmospheric pressure 80kPa~105kPa
Battery
4 "AA" alkaline batteries, AC Adapter (AC 220V) separately sold
Dimensions
130*110*80mm
Main Unit Weight 300gram
Anti-electroshock
type
II class BF digital device