TESTING
TESTING
39
38
10:30 AM | 10/13
121
mg
dL
Control Test
Control Solution Testing
Control Solution Testing
WARNING
• If the control solution test result is out of range,
the result may be transmitted to your pump
when in “Always” send mode (see page 26).
• Do not send an out of range control result to
your pump.
11. Compare your control test result to
the ranges printed on the test strip
bottle.
12. Remove test strip. To turn the
meter off, press and hold the
Menubutton, OR, after three
minutes of inactivity, the meter
turns off. Dispose of the used
test strip as medical waste or
as advised by your health care
professional.
NOTE: Control test results that are marked as
“Control Test” are not transmitted to the pump.
10. The meter shows the
AutoLog screen (if
AutoLog is ON) but will
sense control solution.
The meter will count
down for 5 seconds and the control test result will
display on the meter. It automatically marks the result
as a “Control Test” and stores it in memory. Control
test results are not included in your blood sugar
averages.
CAUTION: If your control solution test result is out
of range, contact Customer Service. See the back
cover for contact information. Do not use the meter
or test strips for blood glucose testing until the
issue is resolved.