Used Test Strip
Remove used strip.
Repeat test with new
strip.
E02
Error Detection Displays
An error screen will always have an “E” with a number
in the lower left corner of the display.
If there is an error (hardware, software, testing errors)
detected by your meter, your meter will beep twice.
You will see specific instructions to guide you. The first
line of the error screen will tell you what the error is.
The next lines describe what you should do. When an
error screen has an OK shown, press the button next to
it to continue.
NOTE: If your screen displays a “connection
error,” your test result was not sent to your
compatible MiniMed insulin pump. Make sure your
Contour
®
next LINK meter is within 4 feet of your
compatible MiniMed insulin pump and the device is
set to accept meter results. Your meter ID must be
programmed into the device. Refer to the separate
Medtronic User Guide.
Check the device display to see if your meter’s
glucose result has been received. If it has not, you can
enter the test result manually. If you continue to have
problems, contact Customer Service. See the back
cover for contact information.
Error Detection Displays
Transferring Data to Medtronic’s CareLink
®
Personal Therapy Management Software
IMPORTANT:
The Contour
®
next LINK meter has only
been tested for use with GLuCofaCts
®
DeLuxe
diabetes management software and
Medtronic’s Carelink
®
Personal Therapy
management software. Ascensia DiabetesCare
is not responsible for any erroneous results
from the use of other software.
For more information, contact Customer Service.
See the back cover for contact information.