System Specifications
Test Sample: Capillary whole blood
Test Result: Referenced to plasma/serum glucose
Sample Volume: 0.6 µL
Measuring Range: 20–600 mg/dL
Countdown Time: 5 seconds
Memory: Stores most recent 1000 test results
Battery Type: Non-serviceable, 310mAh rechargeable
lithium polymer battery, 3.4v – 4.2v (5V input voltage)
Meter/Battery Life: 5 years
Charging Current: 500 mA
Strip Storage Temperature Range: 41º – 86ºF
Strip Humidity Storage Range: 10% – 80% Relative
Humidity (RH)
Control Storage Temperature Range: 48º – 86ºF
Control Solution Operating Temperature
Range: 59ºF – 95ºF
Meter Operating Temperature Range: 41º – 113ºF
Meter Operating Humidity Range: 10% – 93% RH
Dimensions: 3.8 in (H) x 1.2 in (W) x 0.8 in (T)
Weight: 48 grams
Sound: A beep sounds when the meter is turned
on or off, a test strip is inserted, when a test strip is
filled with blood, or when a test result appears on the
screen. Two beeps sound when a strip is removed
and the meter is turned off, or to alert the user. Twenty
beeps will sound when a programmed Reminder
sounds.
Radio Transmitter Frequency: 916.5 MHz
Maximum Radio Transmitter Power: 0 dBi EIRP
Electromagnetic Compatibility (EMC): The
Contour
®
next LINK meter complies with
the electromagnetic requirement specified in
ISO15197:2013. Electromagnetic emissions are
low and unlikely to interfere with other nearby
electronic equipment, nor are emissions from nearby
electronic equipment likely to interfere with the
Contour
®
nextLINK meter. The Contour next LInK
meter meets the requirements of IEC61000-4-2 for
immunity to electrostatic discharge. Avoid use of
electronic devices in very dry environments, especially if
synthetic materials are present. The Contour next LInK
meter meets the requirements of IEC61326-1 for
radio frequency interference. To avoid radio frequency
interference, do not use the Contour next LInK meter
near electrical or electronic equipment that are sources of
electromagnetic radiation, as these may interfere with the
proper operation of the meter.
Caring for Your Meter
• CAUTION: Only connect to a computer or charger
that is certified to UL 60950-1, 5V maximum.
A charging current of 500 mA minimum is required.
• If you use a cable, use the one provided in your
meter kit.
• Store your meter in the carrying case provided
whenever possible.
• Wash hands and dry them well before handling to
keep the meter and test strips free of oils and other
contaminants.
• Handle the meter carefully to avoid damaging the
electronics or causing other malfunctions.
• Avoid exposing meter and test strips to excessive
humidity, heat, cold, dust, or dirt.
Caring for Your Meter
System Specifications