Kangaroo™ Connect Enteral Feeding Pump | 22
Table of Contents
Section 10 — Specifications
Specifications
Medical equipment
Kangaroo Connect enteral feeding pump
(1) Classified with respect to electrical shock, fire, and
mechanical hazards in accordance with ES60601-1 (3rd
edition) and UL60601-1 (2nd edition).
(2) Classified with respect to electrical shock, fire,
mechanical and other specified hazards in accordance
with CAN/CSA C22.2 No. 601.1
Type Infusion Device
Volumetric Enteral
Pumping Mechanism
Rotary Peristaltic
Service Life
The pump, WCH, WiFi Module and accessories are designed
to provide a minimum of 5 years of service life
Shelf Life
The pump, WCH, WiFi Module and accessories are designed
to provide a minimum of 5 years of shelf life
Feeding Sets
All feeding sets are designed to operate only with the
Kangaroo Connect enteral feeding pump
Feeding Formula Delivery Rate
1-600 mL/hr in 1 mL increments
Priming Rate
1310 ± 75 mL/hr
Feeding Formula Dose
1 - 3000 mL in 1 mL increments
Bolus Volume
1 - 3000 mL in 1 mL increments
Number of Boluses
1 - 99
Bolus Interval
1 - 24 hours in 1-hour increments
Accuracy
5% or 0.5 mL/hr, whichever is larger, for all delivery rates no
matter the type of Kangaroo Connect feeding set.
The top of the fluid column should be at a starting height
of 25.4 cm (10”) ± 0.76 cm (0.3”) above the top of the pump.
Accuracy testing is run at a room temperature of 22°C ± 2°C (72°F ± 3°F), using a
new Kangaroo Connect feeding set for no longer than the recommended hours
of usage. Confirmation of accuracy is conducted per the IEC 60601-2-24 standard
for Infusion Devices, as applicable for enteral feeding. For more information on
pump accuracy, see Appendix A.
Occlusion Pressure
Maximum Occlusion Pressure: 20 psi (138 kPa)
Dimensions
Pump: Height: 9.9 cm (3.9”) Width: 15.4 cm (6.1”)
Depth: 4 cm (1.6”)
WCH: Height: 12.8 cm (5”) Width: 19.2 cm (7.6”)
Depth: 9 cm (3.5”)
Weight
0.33 kg (.73 lbs), 0.635 kg (1.4 lbs) with pole clamp
Material
Soft-Touch Coating: Latex-Free Thermoplastic Urethane
Pump Housing: Flame resistant Polyester/Polycarbonate blend
WCH Housing: Flame resistant ABS/Polycarbonate blend
High Priority Alarm Volume
Minimum of 65 dBA at 1 meter in maximum volume orientation
Medium Priority Alarm Volume
Minimum of 60 dBA at 1 meter in maximum volume orientation
Operating Temperature for All System Components
5° - 40° C (41° - 104° F)
Operating Humidity for All System Components
15% - 93% R.H. non-condensing
Packaged Storage and Transport Temperature for All System Components
0° - 50° C (32° - 122° F) 93% R.H. non-condensing
Unpackaged Storage and Transport Temperature for All
System Components
Store between 0º – 50º C (32º F- 122º F) at < 93% RH (non-condensing);
Excursions permitted to -25º C for up to 24 hours.
The pump requires 25 minutes to warm up to its normal operating temperature
from low storage temperatures.
The pump requires 6 minutes to cool down to its normal operating temperature
from high storage temperatures.
Do not operate until the pump is within its operating temperature range.
Ambient Air Pressure
Operating atmospheric pressure range from 62 kPA to 106 kPA
Maximum Altitude
The maximum altitude for using the pump and wireless accessories is 4000m.
Be sure to meet ambient air pressure and battery charging limits stated in
this manual.

Other manuals for Covidien Kangaroo