Section 15 — Electromagnetic Conformity Declaration
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Kangaroo Connect pump, WCH and WiFi module are intended for use in the electromagnetic environment specified below.
The user of the Kangaroo Connect pump, WCH and WiFi module should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment -Guidance
Electrostatic discharge (ESD)
(EN 61000-4-2 per
EN 60601-1-2: 2007)
± 8 kV contact
± 15 kV air
Pump and WCH
± 8 kV contact
± 15 kV air
WiFi Module
± 8 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines ±2 kV for power
supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode ± 1 kV differential mode Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Nominal Mains Voltage
(V
NOM
) 100 VAC and 240 VAC
@ 50 Hz
Voltage Dips
100% of V
NOM
for 10 mSec (0.5
Line Cycles) at 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
100% of V
NOM
for 20 mSec (1
Line Cycle) at 0°
30% of V
NOM
for 500 mSec (25
Line Cycles) at 0°
Interruptions
100% of V
NOMº
for 5000 mSec
(250 Line Cycles)
No degradation of
performance or loss
of function.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Kangaroo Connect pump,
WCH and WiFi module require continued
operation during power mains interruptions, it
is recommended that the device be powered
from an uninterruptible power supply or
a battery.
Power frequency (50/60 Hz)
magnetic field
(EN 61000-4-8 per EN 60601-1-
2: 2007)
30 A/m 30 A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.