C100 Manual
Page 47 - Revision:EU-2020-10-13
4.
Compressor is Not Running
Possible Reason
Solution
Red Heater and Compressor
Switch button on unit is in the
off position
Turn Red Heater and Compressor Switch on.
I = ON
Compressor Starting Circuit
Turn Red Heater and Compressor Switch off. O =
OFF.
Remove the Compressor cap on side of the
Compressor;
Disconnect the black and red terminal connectors;
Inspect the starter and overload relay for any defects.
Replace components(s) as needed.
Turn Red Heater and Compressor Switch on I = ON
and retest Compressor operation.