Spyder 3 CL User's Manual PRELIMINARY 95
DALSA 03-032-20008-03
Message Description
Error 04: Incorrect
parameter value>
This response returned for
· Alpha received for numeric or vice versa
· Not an element of the set of possible values. E.g., Baud Rate
· Outside the range limit
Error 05: Command
unavailable in this
mode>
Command is valid at this level of access, but not effective. Eg line rate
when in smart Exsync mode
Error 06: Timeout> Command not completed in time. Eg FPN/PRNU calculation when no
external Exsync is present.
Error 07: Camera
settings not saved>
Tried saving camera settings (rfs/rus) but they cannot be saved.
Error 08: Unable to
calibrate - tap
outside ROI>
Cannot calibrate a tap that is not part of the region of interest.
Error 09: The
camera's
temperature exceeds
the specified
operating range>
Indicates that the camera has shut itself down to prevent damage from
further overheating.