Gain Accuracy: ± 0.02% of full scale,
typical
, following calibration*
Gain Stability:
vs. Temperature : ± 50 ppm / °C.
vs. Time: ± 20 ppm / month.
Filter: 3-pole modified Butterworth; 3 dB down at 20 Hz; 60 dB down at
220 Hz.
Step Response Settling Time (Full-Scale Output):
To 1% of final value: 60 milliseconds.
To 0.1% of final value: 80 milliseconds.
To 0.02% of final value: 125 milliseconds.
Analog Peak Memory Decay Rate: Does not exceed 0.4% of full
scale/second (e.g., with full scale of 20000 counts, decay rate will
not exceed 8 counts/100 msec).
Analog Output: ±10 V, microprocessor driven and scaled, ±1 mV
resolution.
Accuracy: 0.05% of current voltage reading ± 2 mV.
Output Bandwidth: 40 Hz max.
1.a Model 4077 Description and Specifications
1.5
* Initial (uncalibrated) inaccuracy may be as great as ±0.05% of full scale.
Maximum error that could occur upon replacement of a Model 4077
not
followed by calibration
is ±0.1% of full scale.
Maximum allowable count tolerance is 0 ± 5 (for ZERO) and 20000 ± 11 (for
SPAN, where "m" = 20000).