The PosiTector SST can be adjusted to a known conductivity standard. Use the (-)(+) buttons to
decrease/increase the displayed value to match the known conductivity of the standard.
NOTE: In typical use, a calibration adjustment should not be required. Before performing the
adjustment, ensure that the PosiTector SST test cell has been thoroughly cleaned with DI water.
Ensure that the conductivity standard and PosiTector SST test cell are at the same temperature.
Set the volume of water being used for the test (3.0ml default). Use the (-)(+) buttons to
decrease/increase the displayed value. The patch volume is used to calculate surface density.
When checked (default), the Gage guides the user through a 3-step test process in accordance
with ISO 8502-6. Uncheck Bresle Mode to use the PosiTector SST as a regular conductivity
meter such as when testing abrasive materials in accordance with ISO 11127-6 and ASTM
D4940-15
.
Inverses the LCD display to black on a white background to provide better readability in some
surroundings.
Toggle the Gage sounds ON (checked) or OFF (unchecked).
This option causes the display to read upside down.
Selects display brightness (Sun, Normal or Night). All settings will dim slightly after a period of no
activity to conserve battery life. Press the Down button to brighten the display.
All batches are date-stamped when created, and all readings are time-stamped (24 hour format)
when stored into memory. It is therefore important to keep both the date and time current using
this menu option. Use the Up and Down buttons to scroll, and the (-) and (+) buttons to adjust a
value. The instrument’s data and time can also be viewed in Info.
Converts displayed and printed words to the selected language.