WARNING
Risk of injury or damage
• When the device is operated at the extremes of the environmental operating specications (i.e., maximum temperatures and humidity), and in a single fault condition,
which is a single component or performance malfunction, such as a blocked exhaust vent, blocked air intake, or an internal cooling fan failure, the temperature of the air
coming out of the exhaust vents, located on the bottom left and bottom right sides of the unit can reach temperatures capable of causing a burn injury (see stated
temperature and contact time values in the model specic tables below).
• Keep exposed body parts, such as hands and feet, a minimum of 46 inches (1.2 meters) away from the exhaust vents to avoid the risk of burns. Single fault conditions
may result in visual and audible alerts and alarms.
NOTE – Under normal and single fault conditions, the concentrator releases warm air out the bottom of the unit (exhaust vents) which may discolor temperature
sensitive ooring surfaces. The concentrator should not be used over ooring that is sensitive to heat staining. The Manufacturer is not responsible for ooring that
becomes discolored.
• Operate the unit in a cool, dry area with good ventilation, located on a hard surface, avoid thick rugs or carpeting. NEVER block the air intake or exhaust vents. Keep
the unit a minimum of 12 inches (30.5 cm) away from any wall, draperies, or any other objects that might prevent the proper ow of air in and out of your oxygen
concentrator. Proper air ow is needed to prevent overheating of the oxygen concentrator. DO NOT place the concentrator near any heat source such as hot air
registers or heaters. Overheating of the oxygen concentrator may lead to low oxygen output and a risk of burns .
• The oxygen concentrator should be located in a well-ventilated area. DO NOT operate the unit in a closed or conned space, such as a closet, bathroom, etc. Avoid
operating the device near smoke pollutants and fumes.
• Under extreme environmental conditions and a single fault condition occurs, the following device surface temperatures may exceed 106 °F (41 °C). See Table 1 below
for the model specic maximum temperature and safe contact guidance:
Table 1 - 525DS Series
Description
Maximum Temperature
Max safe contact time
525DS Series
°F °C
Air coming from exhaust vents located on each side
near the bottom of the unit
147.0 63.9 Less than 1 minute
Oxygen outlet tting 120.7 49.3 Less than 10 minutes
Power Switch 119.0 48.3 Less than 10 minutes
LED Indicator panel 129.5 54.2 Less than 1 minute
Cannula at the outlet 109.1 42.8 Less than 10 minutes
Table 2 - 525KS Series
Description
Maximum Temperature
Max safe contact time
525KS Series
°F °C
Air coming from exhaust vents located on each side
near the bottom of the unit
169.0 76.1 Less than 10 seconds
LED Indicator panel 110.5 43.6 Less than 10 minutes
Table 3 - 525PS Series
Description
Maximum Temperature
Max safe contact time
525PS Series
°F °C
Air coming from exhaust vents located on each side
near the bottom of the unit
142.2 61.2 Less than 1 minute
Power Switch 106.7 41.5 Less than 10 minutes
LED Indicator panel 108.5 42.5 Less than 10 minutes
CAUTION
• Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
• It is very important to follow your oxygen prescription. Do not increase or decrease the ow of oxygen – consult your physician.
• Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid inadvertent
damage.
IMPORTANT
• It is recommended that the homecare provider lock the ow control knob to prevent inadvertent adjustment. A ow setting other than prescribed may affect the patient
therapy.
• Do not service or clean this device while in use with a Patient.
• Installation of 515LF-607 low output ow meter package will cause the low ow alarm to not work and will prevent the device from meeting the requirements of ISO-
80601-2-69:2014 Section 201.13.2.101.
• The Device is classied as IP21 which means it is protected against nger access to hazardous parts and protected against vertically falling water drops.
• Equipment not suitable for use in the presence of a ammable anesthetic mixture with air or with oxygen or nitrous oxide.
• This device contains electrical and/or electronic equipment. Follow local governing ordinances and recycling plans regarding disposal of device components.