126 | Dexcom G4 Continuous Glucose Monitoring System
Emissions Test Compliance Electromagnetic Environment
Guidance
RF Radiated emissions
CISPR 11
Group 1,
Class B
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30%.
RF Conducted emissions
CISPR 11
Class B The Dexcom G4 System is suitable
for use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuations/
icker emissions
IEC 61000-3-3
Not applicable
Guidance and Manufacturers Declaration - Electromagnetic Emissions
The Dexcom G4 System is intended for use in the electromagnetic environment
specied below. The customer or the user of the Dexcom G4 System should
ensure that it is used in such an environment.