Dexcom ONE User Guide
77
Professional use instructions
Follow the instructions to clean and disinfect the
receiver and transmitter between patients.
1. Prepare receiver
a. Charge and reset receiver:
Resetting the receiver removes the previous
patients data. To ensure patient privacy, reset
the receiver after each use.
b. Clean and disinfect
See instructions in the Storage and
Maintenance section in Appendix C.
2. Clean transmitter
a. Prepare
Protect: Wear clean gloves and goggles
Prepare Soak: Put Clorox Healthcare
®
Bleach Germicidal Cleaner solution (Clorox)
in a container deep enough to submerge the
transmitter
b. Clean
Rinse and Scrub: Rinse transmitter in cold tap
water while brushing with a soft bristle brush
until all visible soil is gone
Soak and Scrub:
Put transmitter in prepared soak for 3
minutes
While immersed, scrub uneven areas (see
green arrows) with a soft bristled brush or
a pre-saturated bleach wipe