1.4.2 TEST METHODS
Following test are provided to detect of the human coagulation system, which can
be bleeding or thrombosis and the monitoring of anti-coagulation drugs like
Heparin or Marcumar.
1.4.3 SPECIMEN COLLECTION
Type: Human citrated plasma
Collection: Veinvein puncture, 1:10 mixed sodium citrate 3.2%
(0.105M)
Centrifugation: 10min at 1500g
Storage: Max 4h after collection at room temperature
Bilirubin: < 50mg/dl
Hemoglobin: < 9000mg/l
Triglyceride: < 2500g/l
In case of differences to the boxinserts of the reagent, always follow the
instructions on the box insert.
Test Name Specimen Method
DIAcheck series
C1
C2
C4
PTB
Prothrombin Time
blood
clot
Yes
No
No
PT
Prothrombin Time
plasma
clot
Yes
Yes
Yes
APTT
Activated Partial
Prothrombin Time
plasma clot Yes Yes Yes
FIB
Fibrinogen
plasma
clot
Yes
Yes
Yes
TT
Thrombin Time
plasma
clot
Yes
Yes
Yes
AT
Antithrombin
plasma
chromogen
No
Yes
Yes
DD
D-Dimer
plasma
immuno
Yes
Yes
Yes
PC
Protein C
plasma
chromogen
No
Yes
Yes
Factors
Factors II, V, VII, VIII,
IX, X, XI, XII
plasma clot No Yes Yes
AF2
anti IIa
plasma
chromogen
No
Yes
Yes
AF10
anti Xa
plasma
chromogen
No
Yes
Yes
HEP
Heparin
plasma
chromogen
No
Yes
Yes
LMH
Low molecular
heparin
plasma chromogen No Yes Yes
APIX
Apixaban
plasma
chromogen
No
Yes
Yes
RIVX
Rivaroxaban
plasma
chromogen
No
Yes
Yes
EDOX
Edoxaban
plasma
chromogen
No
Yes
Yes
DABI
Dabigatran
plasma
chromogen
No
Yes
Yes

Table of Contents