6.2 OPTIC CHECK
Call: Homescreen / green or red LED / Optic
FIGURE 24: OPTIC CHECK
Display and change intensity of LED
Set all LED to max. intensity (42mA)
Print system report (see next chapter)
Reset all channels and re-calibrate optic
Informations on screen Fault condition Troubleshoot
optic signal, when LED is off
optic signal, when LED is on
Remove cuvette and touch "RESET"
Remove cuvette and touch "RESET"
signal of temperature sensor
adjust temperatur or replace sensor