Operator's Manual DIAcheck series Rev-4 Page 43
6.2 OPTIC CHECK
Call: Homescreen / green or red LED / Optic
FIGURE 24: OPTIC CHECK
Button
Caption
Use Function
(1)-(4)
OPTIC xx
Reset QC value
(5)
I=mA
Display and change intensity of LED
(6)
OK
Return to homescreen
(7)
MAX
Set all LED to max. intensity (42mA)
(8)
Print
Print system report (see next chapter)
(9)
+ / -
Change temperature
(10)
Reset
Reset all channels and re-calibrate optic
Informations on screen Fault condition Troubleshoot
LO
optic signal, when LED is off
> 2900
Replace optic board
HI
optic signal, when LED is on
< 28000
Remove cuvette and touch "RESET"
QC
noise of optic signal
> 30
Touch button "OPTIC"
mA
power of LED (intensity)
not [3 - 12mA}
Remove cuvette and touch "RESET"
TMP
temperature in °C
not [36.0 - 38.0°C]
wait 15min
REF
signal of temperature sensor
not [48000 - 52000]
adjust temperatur or replace sensor