Operator's Manual DIAcheck series Rev-4 Page 44
6.3 SYSTEM REPORT
Call: Homescreen / green or red LED / Print
Fault condition are described in chapter “optic check”
SYSTEM REPORT
22.08.2017
System: DIAcheck
Version: V3.01.45
SIN : 03040 01234
PIN: 12345 67890
TEMP: 37.0°C
50981 (target=50992)
Optic:
Lo Hi mA Qc
---------------------------------------------
1:2698 28822 5 6 OK
2:2698 29822 6 3 OK
3:2698 30822 7 1 OK
4:2698 29822 6 0 OK
PT= 26
aPTT= 8
FIB= 17
DD= 0
AT= 0
TOTAL 101
Date of report
name of system
software version
system ident number
product ident number
temperature of optic and digital value
of thermosensor
Optical values
Lo= LED off
Hi= LED on
mA= LED power
Qc= noise of optic
OK= no fault
!!= fault condition
count of performed tests

Table of Contents