Operator's Manual DIAcheck series Rev-4 Page 22
2.4 SWITCHING ON AND OFF
Switching on
Connect with power supply
Important Information:
The instrument requires about 15 minutes to heat up the optic block to 37°C. Afterwards
it is ready for measurement, indicated by a green dot in the top right corner of the display.
If the status symbol does not turn green, even after wating for 25 minutes, press the status
symbol to see the device status to identify the problem.
Switching off
The device supports no power switch. It must be disconnected from power. To do this,
unplug the power adapter from the socket on the device first and then disconnect the
power supply.
Standby
The system switches to standby after 2 minutes of idle operation. In standby mode, the
display brightness is reduced to save display life time and reduce power consumption. The
next touch anywhere on the display disables the standby mode.
Sleep
Open menu and touch the “sleep” button:
The menu bar is displayed on top of screen and only available, if no measurement is
ongoing. The power consumption during sleep is 0,2W.
Wakeup
To wake the device up from sleep, touch the display.
The system can be disconnected under any operation situations. There is no risk of
system damage