EasyManua.ls Logo

Diamond SmartLyte Plus - Page 56

Diamond SmartLyte Plus
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Troubleshooting
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 56 of 58
Message
Action
System Busy
Instrument is performing an active
task and cannot start another task in
parallel.
Wait for a few minutes for active task to finish
and try again
If still not cleared, reboot the analyzer,
MAIN
OPERATIONS MENU>>>INSTRUMENT
SETTINGS>>>SYSTEM>>>REBOOT
Valve Over Temp
The solenoids overheated or are
improperly connected. The solenoid
valve have thermal fuses that reset.
Turn power off for 1 hour to cool.
Check room temperature within specified limits of
15
o
C to 32
o
C
Ensure all pinch valves properly connected.
↑↑↑↑ ↓↓↓↓
If the analyzer displays multiple
arrows up or down instead of the
sample results, the concentration of
the sample is outside of the
measurement range
For urine sample, arrows up instead of the K+-
result indicate that further dilution is necessary
(see chapter 3.3.4: “Urine samples”).
Check for Proper sample preparation (see
chapter 3).
Check for small air bubbles in the sample after
aspiration into the sample chamber.
Check for proper aspiration of standard A.
↑ ↓
Indicates if measured value is above
or below expected values for
temperature, cal values, or sample
values
Temperature sensor is in the sample sensor. Verify
sample sensor plugged in securely
Check room temperature within specified limits of
15
o
C to 32
o
C.
Check temperature test MAIN OPERATIONS
MENU>>>DIAGNOSTICS>>>CHECK
TEMPERATURE
If temperature is not within 1 to 2 degrees of room
temperature, replace sample sensor.
If new sample sensor does not fix the error, replace
CPU
Communication
Error
Bad connection between display,
CPU and interface PC
Check the flex ribbon connectors between
Display, CPU and interface PC.
Restart the unit and ensure pump turns on at
power cycle.
Contact support if problem persists
Drift Error
Old electrode that are drifting.
Sample STD-A readings where not
close to STD-A reading during
calibration
Check the STD-A mV for sample and calibration
Re-calibrate the instrument
Change the electrode
Flash Drive Error
Flash memory storage malfunctioned
Contact support to get CPU replacement
Missing Data
Or Invalid Data
between interface PC and CPU
board.
Check the flex ribbon connectors between
Display, CPU and interface PC.
Restart the unit and ensure pump turns on at
power cycle.
Contact support if problem persists
No Data Found
Or No Error
display/print/analyze
Check if data is saved properly after each
sample analysis
Check input range and date range
Try to search data remotely using remote
access feature
No Pack Installed
properly
Ensure pack is installed and inserted all the way
in connecting to manifold
Check fluid pack sensor under diagnostic menu
Not Calibrated
running sample
Calibrate the instrument
Ensure current electrode configuration is marked
green on home screen

Related product manuals