1 Introduction
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 8 of 58
1.1. General Description
1.1.1. Measurement
SampleTypes: Whole Blood, Serum, Plasma, QC Material, Aqueous
Reported Parameters Range Measurement Resolution
Sodium, Na
+
40 - 200 mmol/L 0.1 mmol/L
Potassium, K
+
1.5 -15 mmol/L 0.01 mmol/L
Chloride, Cl
-
50 - 200 mmol/L 0.1 mmol/L
Calcium, Ca
++
0.3 - 5.0 mmol/L 0.01 mmol/L
Lithium, Li
+
0.2 – 5.5 mmol/L 0.001 mmol/L
Sample Type: Urine
Reported Parameters Range Measurement Resolution
Sodium, Na
+
1 - 300 mmol/L 1 mmol/L
Potassium, K
+
5 - 120 mmol/L 0.1 mmol/L
Chloride, Cl
-
50 - 200 mmol/L 1 mmol/L
Calcium and Lithium are not measured in urine samples.
1.1.2. Operating Configurations
Sample type Whole blood, serum, plasma, urine, QC material, Aqueous samples
Sample device Syringe, sample cup, collection tube
Sample size 95 uL
Analysis time 28 seconds
Sample rate 128 per hour
Sodium (Na+) sensor Ion-selective, flow-through, glass capillary electrode
Potassium (K+) sensor Ion-selective, flow-through, liquid membrane electrode
Chloride (Cl-) sensor Ion-selective, flow-through, liquid membrane electrode
Calcium (Ca++) sensor Ion-selective, flow-through, liquid membrane electrode
Lithium (Li+) sensor Ion-selective, flow-through, liquid membrane electrode
Reference System Open liquid junction calomel electrode
Calibration Fully automatic
1 point with each sample
2 point every 4 hrs (Lithium, 3 point)
Warm-up time 2 minutes
Temperature Room temperature, 15 - 32° C, 60 - 90° F
Humidity Maximum 85% relative humidity, non-condensing