Operators Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 10
-
2. Press ”over 0.5 second to turn on the machine. The AC power ”indicator light will
illuminate and the battery indicator “ ” will flash and the charge will begin;
3. Press over 0.5 second to turn off the power. The “ flashing indicates it is in charging
state. Approximately 16 hours later the indicator light stop flashing, indicating the battery
is fully charged.
3.2 Installation
3.2.1 Install the machine
1. To attach the pump firmly on the IV pole, please use the knob bolt and pole holder located on the back
panel. Place IV pole through the opening and tight knob bolt firmly.
2. Connect power cord to receptacle, turn on the switch in the rear panel, press ”over 0.5 second.
If “ ” flashes it indicates that the power supply is connected.
3. The infusion pump will perform self-check automatically after it is turned on. This test includes: if the
battery is charged, if the drop detector is attached, if all the system functions are normal. If something
is not correct, an alarm will be trigged indicating the error code (such as AIR, LOW BATTERY, Errs
etc.).
4. After turning on, the device will check the battery automatically. If the charge is low, it will recharge
automatically and “ ” will flash.
Note: If the machine is not connected to the AC power or if the power is cut off, the pump will
change automatically to battery operation and the indicator will come on. The indicator
”flashes and sound alarm is triggered indicating AC power loss. Press any key to clear the alarm.
3.2.2 IV Set Installation:
1. Close of IV set roller clamp. Insert IV set drop chamber sharp extremity into the outlet of the solution
container .
2. Press drop chamber of IV Set gently between fingers
few times to fill 1/4-1/3 chamber full ,as shown in the picture.
Turn the roller clamp
left to close it