Operator’s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 4 -
Chapter I. Overview
1.1. Application
The Digipump LP8 infusion pump is intended to be used in health care facilities for continuous
and accurate infusion of liquids to patients.
1.2 Technical Specification
Iv set requirement 10, 15, 20, 60drops/ml (standard PVC with outer diameter: 3.4mm-4.5mm
(0.134” – 0.178”))
Pump type Linear Peristaltic pump
Control method Drop count method
Air detection method Ultrasound (8 levels of sensitivity are available)
Drop detection method Infrared photo-electric beam
KVO (keep vein open) 1ml/hr (the lowest infusion rate to keep vein open)
Maximum flow rate 999.9ml/h - 10,15, 20 drops/ml IV set
Drop rate accuracy ±2% (the specified accuracy is reached after the first 150 drops).
Flow rate accuracy ±5% (the specified accuracy is reached using adjusted IV set and after
continued infusion over 100ml).
Occlusion pressure 40~140Kpa. Eight levels of sensitivity available
Volume infused display 0-9999ml
Volume to be infused setting 1-9999ml
Sound and light alarm Air in line, door open, occlusion, pause overtime, infusion complete,
abnormal drops, leakage, tube slope, low battery, damage battery, pump
reversal, empty, abnormal drop detector, linear pump cassette abnormal,
sound abnormal, low volume for adjustment, adjustment failure, adjustment
function finished and cannot restart again, interruption of power supply.
Power source ≤25VA
Peristaltic linear pump Detachable
Tube clamp Detachable
Display Color LCD with bright backlight