Operators Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc.
- 25
-
hours,“ ”stop flashing and the battery will be fully charged.
4.2.2 Replacement
1) Loosen screw to remove battery cover. Unplug connector and remove battery.
2) Replace battery, and attach battery cover. Make sure that the battery cover does not smash the
battery cable.
4.3 Periodical replacement of some items
The following parts should be replaced periodically, the replacement time may be shortened
depending on conditions.
4.4 Storage
Store the DigiPump LP8 in the following environment after cleaning.
Ambient temperature : +5 °C +40°C ( 41°F 104°F) Relative humidity : ≤90%
Atmospheric pressure : 70KPa~106Kpa
Do not store the DigiPump LP8 in the following places:
a) Direct sun light or strong light.
b) Direct wind from heating apparatus, air conditioner, ventilator or humidifier, etc.
c) Where there is water splash-up.
d) Where chemicals are stored or gases are generated.
e) Where there is excessive dust or vibrations.
f) Where the floor is not level.
Parts Recommended replacement period
Battery 2 years
Finger pump cassette 3 years